Test Connections
The 6330 comes with the following supplied accessories:
| Part Number |
Description |
| 99-10467-01 |
Adapter Box |
| 99-10457-01 |
LLT Case Lead |
| 99-10468-01 |
LLT Ground Lead |
| 99-10009-01 |
Ground Bond Output |
| 99-10008-01 |
Return Lead |
| 99-10469-01 |
LLT Input Line |
| 99-10470-01 |
LLT Input Neutral |
| 99-10471-01 |
LLT Output Line |
| 99-10472-01 |
LLT Output Neutral |
| 102-055-913 |
High Voltage Cable |
In order to perform a Line Leakage and/or Functional Run test, an external power supply must be used to provide power to the device under test. The power supply must be an unbalanced, single phase 0-277 VAC supply.
Line Leakage / Functional Run stand alone test lead connections
To connect the 6330 to your power source, we provide two black input power leads terminated in one end (p/n’s 99-10469-01 & 99-10470-01). Connect the terminated end in the DUT Power input terminals on the back panel of the 6330. The un-terminated ends are to be connected to the power source.
Also provided with the 6330 are two white output power leads terminated in one end (p/n’s 99-10472-01 & 99-10471-01). Connect the terminated ends in the DUT output terminals on the back panel of the 6330. The un-terminated ends are to be connected to your DUT
Connect the LLT Ground lead into the back panel Ground connection terminal of the 6330. Then connect the LLT Case lead to the Case connection terminal of the 6330. Then connection the LLT Ground lead to your test fixture and the LLT Case lead to the chassis of your DUT.

Adapter Box Connections
The 6330 comes standard with an Adapter Box (p/n 99-10467-01) that provides an easy way to connect a DUT that is terminated in a three-prong line cord. Please note that the adapter box may only be connected to the rear panel of the 6330.
To connect the 6330 to your power source, we provide two black input power leads terminated in one end (p/n’s 99-10469-01 & 99-10470-01). Connect the terminated end in the DUT Power input terminals on the back panel of the 6330. The un-terminated ends are to be connected to the power source.
Then connect the DUT output line connection lead from the adapter box to the back panel DUT output line connection terminal of the 6330. Next connect the DUT output neutral connection lead from the adapter box to the back panel DUT output neutral connection terminal of the 6330. Then connect the ground lead from the adapter box into the back panel Ground connection terminal of the 6330. Next connect the LLT Case lead to the Case connection terminal of the 6330. Then connect LLT Case lead to the chassis of your DUT. Last plug the line cord of your DUT into the adapter box.

|