Test Connections
To connect the 4320 to a DUT, first connect the black return lead into the front panel return output jack. If you are performing a ground bond test then connect the high current lead to the front panel current output jack. Next connect the high voltage lead (w/ red clip) or probe, which-ever one you are using for your application into the high voltage jack on the front panel.
Then attach the black return clip to exposed or dead metal located on the chassis of the DUT. Always connect the ground return first.
Again if you are performing ground bond testing, attach the high current lead to the ground/earthing contact. If you are using the high voltage clip, make sure you attached the red clip to the current carrying conductors of the DUTs circuitry. If you are using the test probe, simply apply the probe to the appropriate current carrying conductors.

Adapter box connection
The adapter box is provided as an optional accessory for the 4320. This provides an easy way to connect to a DUT that is terminated in a two-prong or three-prong line cord.
First connect the black return lead into the front panel return output jack. Then connect the high current lead from the adapter box into the front panel output current jack on the 4320. Then connect the high voltage lead from the adapter box into the front panel high voltage jack on the 4320. Next attach the black return clip to exposed dead metal located on the CHASSIS of the DUT. Always connect the ground return first. Finally plug in your DUT into the adapter box.

|