Test Connections
To connect the 2660 to a DUT, first connect the black return test lead to the return output jack on the front panel of the 2660. Then connect the high current lead (w/red clip) into the high current output jack on the front panel.
Next attach the black return clip to exposed or dead metal located on the chassis of the DUT. Always connect the ground return first. Then connect the high current lead to the ground/earthing contact.

2660 & 2900 Series Interconnection
The 2660 and the 2900 series of instruments can be interconnected together to form a complete test system. First connect the interconnect cable (p/n 99-1007-01) to the front panel return jack of the hipot tester to the return output jack on the front panel of the 2660.
You have the choice to perform your hipot and ground bond testing sequentially (ground bond first, then hipot) or simultaneously (ground bond and hipot at the same time).
For sequential testing use the following connections:
Locate the cable marked "99-10003-01 SEQUENTIAL." Attach the male connector (male connector with blue tubing) with two cables coming from it, to the female input connector of the 2660. Attach the shorter of the two remaining cables with female connectors (female connector with blue tubing), to the male output connector of the 2660 and the longer cable to the male signal input/output connector (female connector with red tubing) on the rear panel of the Hipot.
For simultaneous testing, use the following cable connections:
Locate the cable marked "99-10004-01 SIMULTANEOUS." Attach the male connector (male connector with blue tubing) with two cables coming from it, to the female input connector of the 2660. Attach the shorter of the two remaining cables with female connectors (female connector with blue tubing), to the male output connector of the 2660 and the longer cable to the male signal input/output connector (female connector with red tubing) on the rear panel of the Hipot.

After your testers are interconnected you may then connect the system to a DUT. First connect the return lead (p/n 99-10238-01) to the front panel return jack on the 2660. Then connect the high current lead (p/n 99-10239-01) to the high current output jack on the front panel of the 2660. Then connect the high voltage lead (p/n 102-055-913) to the front panel high voltage output jack on the 2900 series instrument.
Then connect the return to exposed or dead metal located on the chassis of the DUT. Always connect the ground return first. Then connect the high current lead to the ground/earthing contact of the DUT. Lastly connect the high voltage lead to the line and neutral contacts of the DUT.

|