Quick Start
The 6330 comes ready with the following factory default settings in memories 1 – 10. These default settings will remain unless overridden with your own test program. Before beginning, please make sure the input voltage selector switch is set correctly on the back panel of the instrument (115 or 230)
Defaults
- Input Voltage: 115 or 230 Volts AC country specific (rear panel switch selectable)
- PLC Remote: Off
- Single Step: Off
- Fail Stop: On
- Lock: Off
- Memory Lock: On
- Alarm Volume: 5
- DUT HV-SET: On
To edit system parameters press exit from the test screen. You can then press the "set" key to toggle through the parameters and the up and down keys to edit the parameters.
Test Parameters
- Memory Position: 1
- Step Position: 1
- Test Mode: AC
- Voltage: 1.24 kV
- HI-Limit: 10.00 mA
- Lo-Limit: 0.00 mA
- Ramp: 0.1 s
- Dwell: 1.0s
- Frequency: 60 Hz
- DUT Output: On
- Connect: Off
Changing the test mode will change the defaults used. The above defaults are for an AC Hipot test mode. Following are the defaults for DC Hipot, Insulation Resistance, Ground Bond, Functional Run, and Line Leakage Testing.
DC Hipot Mode
- Voltage: 1.50 kV
- HI-Limit: 5.00 mA
- Lo-Limit: 0.00 mA
- Ramp: 0.1 s
- Dwell: 1.0s
- DUT Output: On
- Connect: Off
Insulation Resistance
- Voltage: 500V
- HI-Limit: 0 mΩ
- Lo-Limit: 1 mΩ
- Delay: 1.0 s
- DUT Output: On
- Connect: Off
Ground Bond Test
- Current: 30 Amps
- HI-Limit: 100 mΩ
- Lo-Limit: 0 mΩ
- Dwell: 1.0s
- Offset: 0 mΩ
- Frequency: 60 Hz
- Connect: Off
Functional Run
- Volt-Hi: 125.0V
- Volt-Lo: 0.0V
- Amp-Hi: 10.0A
- Amp-Lo: 0.0A
- Power-Hi: 1000W
- Power-Lo: 0W
- PF-Hi: 1.000
- PF-Lo: 0.000
- Leakage-Hi: 10.00 mA
- Leakage-Lo: 0.00 mA
- Delay: 0.2s
- Dwell: 1.0s
- Connect: Off
Line Leakage Mode
- Volt-Hi: 125.0V
- Volt-Lo: 0.0V
- Leakage-Hi: 6000 µA
- Leakage-Lo: 0.00 µA
- Line: 7
- MD: MD A
- Dalay: 1.0s
- Connect: Off
There are seven different Line Conditions to set up. They are displayed in the following table:
| Line |
1 |
2 |
3 |
4 |
5 |
6 |
7 |
8 |
| Neutral |
OPEN |
OPEN |
OPEN |
OPEN |
CLOSE |
CLOSE |
CLOSE |
CLOSE |
| Reverse |
OFF |
ON |
OFF |
ON |
OFF |
ON |
OFF |
ON |
| Ground |
OPEN |
OPEN |
CLOSE |
CLOSE |
OPEN |
OPEN |
CLOSE |
CLOSE |
The 6330 comes equipped with 5 different Measuring Device models. Following is a cross reference:
| Cross Reference Letter |
Agency Specification |
Application |
| A |
UL544NP, UL484, UL923, UL471, UL867, UL697 |
Medical Equipment |
| B |
IEC60990 Fig4 U2, IEC 60950-1, IEC60335-1, IEC60598-1, IEC60065, IEC61010 |
Medical Equipment |
| C |
UL60601-1, IEC60601-1 |
Medical Equipment |
| External |
User configurable |
User configurable |
| Frequency Check |
The Line Leakage test can be configured to verify the bandwidth of the leakage current "voltmeter." |
The selection is intended to be used for meter verification only and has no application for normal testing. |
To edit parameters, press set from the test screen. You will now be able to use the "set" key to toggle through each parameter. The "up" and "down" keys will change the parameter.
|