Page 24 - SCI Catalog.indd
P. 24

INDEX
SAFETY STANDARD REFERENCE CHART
Standard / Testing Dielectric Withstand Harmonized Type
Standard Test Voltage
Ground Bond/Continuity
Earth Leakage
Insulation Suggested Resistance Model #
Max I.
Test Time
Test Current
VLimit
Max. R
Test Time
Test Voltage
Max I.
Test Time
VLimit
Min Slaughter Tester R
IEC/UL 60601-1
3rd Edition Medical Electrical Equipment
Performance
Production
500 – 4000 VAC or 707 – 5656 VDC
1000 – 3000 VAC
No Breakdown
60 s
10-25 A
≤ 6V
≤ 0.1 Ω
5s
110% x rated V
5-10 mA
N/A 6330
1 or 60 s
10-25 A
≤ 6V
≤ 0.1 Ω
5s
N/A
N/A 4320, 4520
H.U.D. Speci cation #24 CFR 3280.810
Performance
Production
900-1079 VAC or 1273-1526 VDC
1080-1250 VAC or 1527-1768 VDC
No Breakdown
60 s
Continuity
N/A
N/A
294, 295, 296, 297
No Breakdown
1s
Continuity
N/A
N/A
294, 295, 296, 297
R.V.I.A. (NEC)
Performance 900 VAC or 1280 VDC
No Breakdown
60 s
Continuity
N/A
N/A
294, 295, 296, 297
Production
1080 VAC or 1530 VDC
No Breakdown
1s
Continuity
N/A
N/A
294, 295, 296, 297
IEC 60335-1 Household Electrical Appliances
Performance
Production
500 – 2400 VAC x rated V + 2400 VAC
400 – 2500 VAC
No Breakdown
60 s
≥ 10 A
≤ 12 V
0.1 – 0.2 Ω
≤ 120 s
1.06 x rated V
0.25 – 5.0 uA
N/A 6330 + 298*
5-30 mA
1s
≥ 10 A
≤ 12 V
0.1 – 0.2 Ω
No time speci ed
N/A
N/A 4520
UL 60335-1 Household Electrical Appliances
Performance
Production
500 V – 2400 VAC x rated V + 2400 VAC
400 – 2500 VAC
No Breakdown
60 s
40 A
≤ 6.5 V
≤ 0.5 Ω
120 s
1.06 x rated V
0.25 – 5.0 uA
N/A 6330 + 298*
5-30 mA
1s
40 A
≤ 12 V
0.1 – 0.2 Ω
No time speci ed
N/A
N/A 4520
IEC 60598-1 Luminaires
Performance
Production
500 – 4 x rated V + 2000 VAC
No Breakdown
60 s
≥ 10 A
≤ 12 V
≤ 0.5 Ω
60 s
Rated V
0.5 – 10 mA
60 s
500 VDC
1-4 6330 + 298* MΩ
Not Speci ed - Responsibility of Manufacturer
294, 295, 296, 297
UL 1598 Luminaires
Performance 1000 VAC - 1000 VAC x 2 x rated V
Production 1200 VAC
No Breakdown
60 s
30 A
≤ 4V
≤ 0.1 Ω
120 s
N/A
No time speci ed
500 VDC
≥ 2 4520 MΩ
1s
Continuity
≤ 0.1 Ω
Continuity
N/A
N/A 295
IEC/UL 61010-1 & CSA 22.2 No. 61010- 1 Laboratory Control Test & Measurement Equipment
Performance
Production
840 - 11940 VAC or 1200 - 7500 VDC
5 - 60 s
25 or 30 A
≤ 10 V or ≤ 12 V
≤ 0.1 Ω or <4 V 0.133 Ω
< 300 V
0.5 mA
N/A 6330 + 298*
No Breakdown
5 s max ramp up 2s dwell
Continuity
60 or 120 s
N/A
N/A
294, 295, 296, 297
EN 60204-1 Electrical Equipment of Machines
Performance
Production
2 x rated V or 1000 VAC
No Breakdown
1s
0.2 - 10 A
≤ 24 V
Refer to Section 18.2.2
No time speci ed
N/A
No time speci ed
500 V
≥ 1 4320 MΩ
Not Speci ed - Responsibility of Manufacturer
294, 295, 296, 297
UL 45A Portable Electrical Appliances
Performance
Production
1000 VAC + 2 x rated V or DC equivalent
1000 - 3000 VAC
EN 60950-1 EN 50116 Information Technology Equipment
Performance
Production or 1414 – 4242 VDC
1000 – 3000 VAC
No Breakdown
No Breakdown
60 s
Continuity
< 300 V
0.5 – 3.5 mA
60 s
500 V
≥ 50 6330 KΩ
1s
Continuity
N/A
N/A 294, 295
120 s
30 A
≤ 12 V
≤ 0.1 Ω
60 s
< 300 V
0.25 – 3.5 mA
60 s
500 V
≥ 2 6330 MΩ
1 -4 s
25 A
≤ 12 V
≤ 0.1 Ω
1-4 s
N/A
N/A 4320
UL 60950-1 CSA 22.2 No. 60950- 1 Information Technology Equipment
Performance Production
1000 – 3000 VAC or 1414 – 4242 VDC
60 s
≤ 40 A
≤ 12 V
≤ 0.1 Ω
60 s
< 300 V
0.25 – 3.5 mA
60 s
500 V
≥ 2 6330 MΩ
No Breakdown
1 –6 s
Continuity
N/A
N/A 294, 295, 296
*This standard requires the use of a 500VA Hipot tester.