  2200 SERIES SPECIFICATIONS
INDEX
    INPUT
   Voltage
  115/230 V selectable, ± 15% variation
 Frequency
  50/60 Hz ± 5%
   Fuse
  1 A 250 VAC fast acting
  GENERAL SPECIFICATIONS
   Remote I/O
 Provided through 9 pin D type connector 1. Inputs: Test, Reset, SafetyInterlock
2. Outputs: Pass, Fail and Test-in-Process
   Calibration
   Software & adjustments made through front panel
   Line Cord
  Detachable 6' (1.80 m) power cable terminated in a three prong grounding plug
  Terminations
  High Voltage Output: Alden Socket
   Shielded Return: BNC Connector
    Dimension (W x H x D)
 4.75" x 5.25" x 11.75" (120 x 133 x 300 mm)
   Weight
   11 lbs. (5 kg)
        INSULATION RESISTANCE TEST MODE
  Output Voltage
   Range: Resolution: Accuracy:
Ripple:
 30 - 1000 VDC
1V
± (1% of setting + 1 V) (relative to displayed output)
< 2%
  Voltage Display
    Low Range: High Range: Resolution: Accuracy:
  0 V - 100 V
101 V - 1000 VDC
0.1 V (low range), 1 V (high range) ± (2% of reading + 2 V)
 Resistance Display
   Range:
   0.01 MΩ - 200.0 GΩ (4 digit, auto ranging)
 Resolution:
      30-499 VDC
   500-1000 VDC
   .001 MΩ -.1MΩ
 .1MΩ-1GΩ
 .1MΩ-1GΩ
   1MΩ -.01GΩ
  1GΩ-20 GΩ
  1GΩ-20GΩ
   0.1GΩ
      20GΩ-200GΩ
    Accuracy:
     30 - 499 V
   0.1MΩ-1GΩ
  ± (3% of reading + 2 counts)
   1 - 20GΩ
  ± (5% of reading + 2 counts)
     500 - 1000 V
   0.1MΩ-1GΩ
  ± (2% of reading + 2 counts)
   1 - 20GΩ
   ± (3% of reading + 2 counts)
   20- 200GΩ
  ± (10% of reading + 2 counts)
   Timer Display
   Range: Resolution: Accuracy:
  0 - 999.99 seconds
0.1 seconds
± (0.1% of reading + 0.05 seconds)
 Failure Settings
  Low Limit:
     0.1 MΩ - 999.9 MΩ
  1000 MΩ - 9999.MΩ
  10.0 GΩ - 200.00 GΩ
    Dwell Timer
  1.0 - 999.9 seconds, 0.1 seconds/step, (0=continuous)
   Delay Timer
  0.1 - 999.9 seconds, 0.1 seconds/step
  Discharge
    Automatic discharge of device under test
  Indicator: Green < 30 V, Red > 30 V
   Why We Use Counts: Slaughter publishes some specifications using “counts” which allows us to provide a better indication of the tester’s capabilities across measurement ranges. A “count” refers to the lowest resolution of the display for a given measurement range. For example, if the resolution for voltage is 1V then 2 counts = 2V.
Specifications subject to change without notice.