2200 SERIES SPECIFICATIONS
INDEX
INPUT
Voltage
115/230 V selectable, ± 15% variation
Frequency
50/60 Hz ± 5%
Fuse
1 A 250 VAC fast acting
GENERAL SPECIFICATIONS
Remote I/O
Provided through 9 pin D type connector 1. Inputs: Test, Reset, SafetyInterlock
2. Outputs: Pass, Fail and Test-in-Process
Calibration
Software & adjustments made through front panel
Line Cord
Detachable 6' (1.80 m) power cable terminated in a three prong grounding plug
Terminations
High Voltage Output: Alden Socket
Shielded Return: BNC Connector
Dimension (W x H x D)
4.75" x 5.25" x 11.75" (120 x 133 x 300 mm)
Weight
11 lbs. (5 kg)
INSULATION RESISTANCE TEST MODE
Output Voltage
Range: Resolution: Accuracy:
Ripple:
30 - 1000 VDC
1V
± (1% of setting + 1 V) (relative to displayed output)
< 2%
Voltage Display
Low Range: High Range: Resolution: Accuracy:
0 V - 100 V
101 V - 1000 VDC
0.1 V (low range), 1 V (high range) ± (2% of reading + 2 V)
Resistance Display
Range:
0.01 MΩ - 200.0 GΩ (4 digit, auto ranging)
Resolution:
30-499 VDC
500-1000 VDC
.001 MΩ -.1MΩ
.1MΩ-1GΩ
.1MΩ-1GΩ
1MΩ -.01GΩ
1GΩ-20 GΩ
1GΩ-20GΩ
0.1GΩ
20GΩ-200GΩ
Accuracy:
30 - 499 V
0.1MΩ-1GΩ
± (3% of reading + 2 counts)
1 - 20GΩ
± (5% of reading + 2 counts)
500 - 1000 V
0.1MΩ-1GΩ
± (2% of reading + 2 counts)
1 - 20GΩ
± (3% of reading + 2 counts)
20- 200GΩ
± (10% of reading + 2 counts)
Timer Display
Range: Resolution: Accuracy:
0 - 999.99 seconds
0.1 seconds
± (0.1% of reading + 0.05 seconds)
Failure Settings
Low Limit:
0.1 MΩ - 999.9 MΩ
1000 MΩ - 9999.MΩ
10.0 GΩ - 200.00 GΩ
Dwell Timer
1.0 - 999.9 seconds, 0.1 seconds/step, (0=continuous)
Delay Timer
0.1 - 999.9 seconds, 0.1 seconds/step
Discharge
Automatic discharge of device under test
Indicator: Green < 30 V, Red > 30 V
Why We Use Counts: Slaughter publishes some speci cations using “counts” which allows us to provide a better indication of the tester’s capabilities across measurement ranges. A “count” refers to the lowest resolution of the display for a given measurement range. For example, if the resolution for voltage is 1V then 2 counts = 2V.
Speci cations subject to change without notice.