Page 22 - SCI Catalog.indd
P. 22

INDEX
6000 SERIES SPECIFICATIONS
INPUT
Voltage Frequency
Fuse
115 / 230 V selectable, ± 10 % variation 50/60 Hz ± 5%
6.3 A slow blow 250 VAC
RUN TEST MODE
DUT Power
Voltage Display
Short Circuit Protection
Delay Timer Dwell Timer Timer Display Voltage Current Watts
Power Factor Leakage Current
0 - 277 VAC Single Phase Unbalanced 30 AAC max continuous
0 - 277.0 VAC Full Scale
Short circuit current 50 A < 3 s Inrush current 180 A Response time 10 μs
0.2 - 999.9 seconds
0, 0.1 - 999.9 seconds (0 = continuous) 0 - 999.9 seconds
0 - 277.0 VAC
0 - 30.0 AAC
0 - 8400 W
0 - 1.000
0 - 10.00 mA (0 = o )
Leakage current measuring resistor MD=2KΩ ± 1%
Voltage: Current:
Range:
DIELECTRIC WITHSTAND TEST MODE
Output Rating Voltage Setting
Current Display
HI-Limit LO-Limit
Failure Detector
DC Output Ripple
Discharge Time
Max. Capacitive Load in DC Mode
AC Waveform
AC Output Frequency Output Regulation Dwell Timer
Ramp Timer
3.5 kV @ 30 mAAC
4.0 kV @ 5 mADC
0 - 3.50 kVAC
0 - 4.00 kVDC
0.01 kV
± (2% of setting + 5 V)
0 - 30.00 mAAC
0 - 5.00 mADC
0.01 mA
± (2% of reading + 0.02 mA)
Range: 0 - 30.00 mA
Accuracy: ± (2% of setting + 0.02 mA)
Range: 0-5.00 mA
Accuracy: ± (2% of setting + 0.02 mA)
Audible and Visual
5% Ripple RMS at 4 kVDC @ 5 mA, resistive load
LEAKAGE CURRENT TEST MODE
Range:
Range:
Range:
Range:
Range:
Range:
Range:
Range:
Range:
Resolution: Accuracy:
Range:
Resolution: Accuracy:
AC
DC
Voltage: Current:
Range:
≤ 200ms
Sine Wave, Crest Factor = 1.3 - 1.5
Range: 50/60 Hz, user selection
< 3 kV < 4 kV
DUT Power
Voltage Display
Short Circuit Protection
Leakage Current (RMS Only)
Measuring Device
Line Condition
0 - 277 VAC Single Phase Unbalanced 0 - 30 A maximum
30 AAC max continuous
0 - 277.0 VAC Full Scale
Short circuit current 50 A < 3s Inrush current 180 A Response time 10 μs
0μA-6000μA
DC, 15 to 100 kHz ± (2% of reading + 3 μA) > 100 k to 1 MHz ± 5% of reading
UL544 Non Patient
IEC60990 Fig4-U2
UL2601-1, UL60601-1
Frequency Check, External MD (1k Ω) External MD (1k Ω)
Neutral, Reverse, Ground
1.00 uF
< 1 kV
0.50 uF
0.75 uF
< 2 kV
0.08 uF
Range:
Accuracy:
A B C F X
± (1% of setting + 5 V) from no load to full load
0, 0.2 - 999.9 sec, (0 = continuous)
0.1 - 999.9 sec
Range:
Range:
INSULATION RESISTANCE TEST MODE
Output Voltage Resistance Display
HI-Limit/LO-Limit Delay Timer
GROUND BOND TEST MODE
Output Voltage Output Frequency Output Current Current Display HI-Limit
LO-Limit
Dwell Timer Milliohm O set
100 - 1000 VDC
± (2% of reading + 5 V)
1 - 1000 MΩ (4 digit, auto ranging) 1000 VDC
MΩ
1.00 - 80.00 75.0 - 999.9
1 - 1000 MΩ (0 = o )
0, 0.5 - 999.9 sec, (0 = continuous)
6 VAC  xed
50/60 Hz, user selectable 3.0 - 30.0 AAC
0.0 - 30.0 A
0 - 510 mΩ for 3.0 - 10.0 A 0 - 200 mΩ for 10.1 - 25.0 A 0 - 150 mΩ for 25.1 - 30.0 A
± (2% of setting + 2 mΩ)
0, 0.5 - 999.9 sec, (0 = continuous)
0 - 100 mΩ
± (2% of setting + 2 mΩ)
Probe  G-L HI-Limit/LO-Limit
Delay Timer
GENERAL SPECIFICATIONS
0 - 6000 μA
0, 1.0 - 999.9 sec, (0 = continuous)
Range: Accuracy:
Range:
500 VDC
MΩ
MΩ
0.01
1.00 - 40.00
0.1
35.0 - 999.9
Range:
Range:
Resolution:
Range:
Memories Remote I/O
Interface
Security
Terminations
Calibration
Mechanical
Environmental
Dimension (W x H x D)
Weight
20 memories with 10 steps per memory
Test, Reset, Interlock & recall memory 1-6 Pass, Fail, Test-In-Process
RS-232 interface
Key lock and memory lock capability to avoid unauthorized access
6' (1.80 m) high voltage and (2) return leads w/ clips
Software & adjustments made through front panel
Bench or rack mount with tilt up front feet
Operating Temperature: 32o - 104o F (0o - 40o C) Relative Humidity: 20 - 80%
16.9" x 5.2" x 19.6" (430 x 133 x 500 mm)
48.5 lbs. (22 kg)
Range:
Input: Output:
Range:
Range:
Range:
Range:
Range:
Accuracy:
Range:
Range: Accuracy
2182
Safety Made Simple® hipot.com