260 SERIES SPECIFICATIONS
INPUT
Voltage
264
100 - 120 VAC / 200 - 240 VAC ± 10% Auto Range
266
100 - 240 VAC ± 10% Full Range
Frequency
50/60 Hz ± 5%
Fuse
264
10A / 250 VAC Slow-Blow
266
12A / 250 VAC Slow-Blow
GENERAL SPECIFICATIONS
Memories
5
Remote I/O
Input:
Test, Reset, Interlock
Output:
Pass, Fail, Test-in-Process
Hardware Interlock - a relay on the high voltage output opens when the Interlock signal is disabled.
Voltage Drop Display (optional)
Display the voltage drop across the circuit instead of the resistance measurement.
Voltage Limit Settings
264
0.00 - 6.00 VAC
266
0.00 - 9.00 VAC
Resolution: 0.01 V
Accuracy: ± (2% of setting + 0.02 V)
O set Limit Settings
264
0.00 - 4.00 VAC
266
0.00 - 6.00 VAC
Resolution: 0.01 V
Accuracy: ± (2% of setting + 0.02 V)
Security
Option to turn On or O , when On you can switch between two security levels:
1. Run - Operator can only run a test. No ability to change memory locations or edit test parameters.
2. Mem - Operator can run a test and change memory locations. No ability to edit test parameters.
Safety Mark
CE/cTUVus
Dimensions
(W x H x D)
264
8.5" x 3.5" x 11.81" (215 x 88 x 300 mm)
266
16.93" x 5.20" x 11.81" (430 x 132 x 300 mm
Weight
264
9.25 lbs. (4.3 Kg)
266
20.25 lbs. (9 Kg)
GROUND BOND TEST MODE
Output Rating
264
3.0 - 40.0 AAC
266
3.0 - 60.0 AAC
Resolution: 0.1 A
Accuracy: ± (2% of setting + 0.1A)
264
Voltage 8 VAC ( xed)
266
Voltage 12 VAC ( xed)
Output Frequency
50/60 Hz user selectable Accuracy: ± 0.1%
Resistance Limit Settings
264
0 - 150 mΩ for 30.1 - 40.0 A 0 - 200 mΩ for 10.1 - 30.0 A 0 - 600 mΩ for 3.0 - 10.0 A
266
0 - 150 mΩ for 30.1 - 60.0 A 0 - 200 mΩ for 15.1 - 30.0 A 0 - 600 mΩ for 3.0 - 15.0 A
Resolution: 1 mΩ
Accuracy: ± (2% of setting + 2 mΩ)
O set Limit Settings
0 - 100 mΩ
Resolution: 1 mΩ
Accuracy: ± (2% of setting + 2 mΩ)
Dwell Timer
0, 0.5 - 240.0 sec, (0=continuous), 0.1 sec/step
Ramp Timer
0.1 sec fixed
Measurement Current
264
0.0 - 40.0 AAC
266
0.0 - 60.0 AAC
Resolution: 0.1 A
Accuracy: ± (3% of reading + 0.1 A)
Ohmmeter
264
0 - 600 mΩ
Resolution: 1 mΩ
Accuracy: ± (3% of reading + 3 mΩ) for 3 - 5.9 A, ± (2% of reading + 2 counts) for 6 - 40A
266
0 - 600 mΩ
Resolution: 1 mΩ
Accuracy: ± (3% of reading + 3 mΩ) for 3 - 5.9 A
± (2% of reading + 2 mΩ) for 6 - 60 A
Speci cations subject to change without notice.